一种经验证的环境友好型高效液相色谱测定法用于猫传染性腹膜炎患猫中GS-441524的治疗药物监测。 机翻

A Validated Eco-Friendly High-Performance Liquid Chromatography Assay for Therapeutic Drug Monitoring of GS-441524 in Cats with Feline Infectious Peritonitis.

中文关键词(术语机标,非人工翻译):猫传染性腹膜炎 · 实验室检查

期刊Pathogens (Basel, Switzerland)
发表日期2026-08-31
物种猫
DOI10.3390/pathogens15090913
PMID42796681

摘要

Feline infectious peritonitis (FIP) is responsive to treatment with the adenosine analogue GS-441524 (GS-44) and its prodrug, remdesivir (REM); both are now available on veterinary prescription in many countries. Therapeutic drug monitoring (TDM) of GS-44 has the potential to support dose selection for individual cats; however, TDM assays are currently only offered by one UK-based laboratory. This study describes a simple, cost-effective, and environmentally conscious high-performance liquid chromatography (HPLC) method for the quantification of GS-44 in feline plasma or serum. The method was validated in accordance with the International Council for Harmonisation M10 guidelines for bioanalytical methods. Calibration standards demonstrated linearity across a range of 2.59 to 5048.28 ng/mL, R<sup>2</sup> = 0.9880, with a lower limit of quantification of 2.59 ng/mL and upper limit of quantification of 5048 ng/mL; this is equivalent to an assay range of 0.06 to 208 µM. Precision, accuracy and spike recovery were within ±15% for 14 of 17 standard concentrations (±20% at the lowest three). Carry-over, dilution integrity, and analyte stability under common storage conditions all met the method's requirements. This is a simple, robust and accurate eco-friendly method suitable for adoption by diagnostic laboratories, enabling routine TDM for cats undergoing treatment for FIP with GS-44 and/or REM.

站内已收录开放获取全文,见下方。

全文 开放获取 · cc by 中文机翻

1. 引言

瑞德西韦(REM)和GS-441524(GS-44)是抗病毒药物,用于治疗患有各种形式猫传染性腹膜炎(FIP)的猫时安全有效[1,2]。目前,在许多国家,这两种药物已可由执业兽医从持证兽药配制公司合法处方获得。

这些药物的治疗方案基于来自体外细胞培养和健康猫研究的数据[3,4,5,6,7,8],以及使用合法来源GS-44和/或REM治疗的猫[4,7,8,9,10,11,12,13,14,15,16,17,18,19,20]或非法来源不可验证化合物治疗的猫[21,22,23,24,25]的回顾性结果报告。迄今为止,在兽医监督下对临床FIP猫给予这些药物的已发表药代动力学(PK)数据很少(n = 6[PK数据],n = 22[单样本数据])[8],需要对适合FIP每种临床表现的剂量进行深入研究[8]。鉴于临床病例可用的预测性PK/药效动力学(PD)数据有限,准确定量GS-44浓度是指导循证给药和治疗监测的重要替代手段。

治疗药物监测(TDM)用于监测目标药物的血浆浓度以优化治疗[26];这主要是为了防止剂量不足或过量,但也用于支持对药物治疗反应失败、副作用或缓解后复发原因的调查。

GS-44的TDM需要一种临床医生易于获得的实验室方法,能够快速(<7天)且经济有效地报告结果。目前,高效液相色谱结合质谱法(HPLC-MS或HPLC-MS/MS[27,28,29,30])是研究人员最常用的方法;然而,该方法使用昂贵且复杂的设备,需要专业技术人员操作,消耗大量环境敏感资源,且(截至2026年6月)尚未提供临床使用。

本文旨在描述一种简单、低成本、有效且不使用MS、仅消耗少量环境敏感资源的HPLC方法的验证,该方法可能在商业诊断实验室中找到应用,并为临床医生提供这一重要资源。

我们还报告了来自361只接受FIP各种临床表现治疗的猫的1034份样本的GS-44检测结果。

2. 材料与方法

2.1. 样本来源与处理

猫血清(自然凝固或血清凝胶)和血浆样本(乙二胺四乙酸[EDTA]和/或肝素抗凝)作为残余诊断材料获得,来源于爱丁堡大学皇家(迪克)兽医学院Easter Bush病理实验室的常规临床送检样本,以及来源于接受合法处方REM或GS-44治疗FIP的猫的诊断样本。使用残余临床样本进行方法开发和验证已获英国爱丁堡大学兽医伦理审查委员会(VERC)批准(VERC 63.22;2022年12月4日批准)。诊断检测后,残余材料被转移至微量采样实验室进行方法开发和验证。

2.2. 化学品、试剂和耗材

所有溶剂(包括超纯水)和化学品均为HPLC级(经0.22 µm过滤),购自英国伦敦LaserChrom HPLC Laboratories,但GS-44(CAS:1191237-69-0,纯度> 99%,在其原始运输容器中于−20 °C储存)除外,其购自瑞典斯德哥尔摩MedChemTronica AB。

2.3. 储备液和标准溶液的制备

使用甲醇(100%)制备36,600 ng/mL的GS-44储备液,于−20 °C储存。用下文所述的HPLC流动相(MP)稀释储备液,制备了17个标准溶液,GS-44浓度范围为5048.28至2.59 ng/mL。甲醇未用作标准品的稀释剂,因为在方法开发过程中,使用甲醇时色谱图基线过度扰动和峰变宽/肩峰现象出现,而使用MP作为稀释剂时未出现这些伪影;这似乎是色谱柱特异性现象。标准溶液在2 mL HPLC样品瓶中制备,于0 °C至+6 °C(实验室冰箱)储存,并在每次检测序列前短暂涡旋。标准溶液与研究样本一起检测,如果其计算浓度偏离预期值超过15%(三个最低标准浓度超过20%),则弃用。根据需要制备新鲜标准品组(大约每6个月)。

2.4. 提取和样本制备

在本文中,术语“样本”指来自EDTA或肝素抗凝血液的血浆或血清。

猫的GS-44血液-血浆分配比被认为为1.2(其他五个物种返回相似值[31]);因此,对于完全溶血的样本,检测值可能比无溶血时获得的检测值高约20%。在本研究中,如果离心无法将红细胞从溶血样本中分离,即发生完全溶血,则这些样本被排除。

导致样品中出现黄疸和脂血现象的分子不会对该方法产生干扰[32,33],因此这些干扰因素未被视作干扰因子。

每个样品制备2个稀释度,使用正压移液器(Gilson Microman;Gilson UK,Dunstable,UK)将220 µL甲醇(通过凝固去除蛋白质)与(a)20 µL样品(1:12稀释)和(b)40 µL样品(1:6.5稀释)混合,以防止粘稠样品导致测量误差。涡旋两次,每次10 s,间隔30 s,以7700 g相对离心力离心6 min后,取210 µL上清液转移至新的1.5 mL小瓶中,在真空条件下蒸干(GeneVac EZ-2,约60 min,≤40 °C)。残渣用210 µL MP重新溶解,使用超声浴(5 min,≤40 °C)并涡旋,然后通过4 mm 0.22 µm尼龙注射器滤器过滤至0.3 mL HPLC小瓶中进行分析。

2.5. 仪器与色谱条件

使用Dionex(Fisher Scientific UK Ltd.,Loughborough,UK)UltiMate RS 3000 UHPLC系统,该系统由四通道MP真空脱气机、温控自动进样器(20 °C)、双泵、柱温箱(色谱柱40 °C;柱后23 °C)和四通道紫外二极管阵列检测器(DAD)组成。仪器控制、数据采集和色谱图分析通过Chromeleon v 7.2(Fisher Scientific UK Ltd.,Loughborough,UK)软件完成。样品进样体积为40 µL,进样间使用MP进行针头清洗。

GS-44分子具有高极性(XlogP3-AA = −1.4 [34]),因此需要高水相含量的MP以充分增加其在C18固相柱上的保留时间,确保与其他样品组分实现足够的峰分离。选择Waters XSelect HSS T3 100Å,2.5 µm,4.6 mm × 150 mm色谱柱(Waters Ltd.,Wilmslow,UK),因为其固相设计可用于含水高达100%的MP,且在方法开发过程中,该柱能使GS-44峰与其他在220至260 nm紫外波长范围内也有吸收的样品组分实现满意分离。洗脱为等度洗脱,MP流速1.5 mL/min,运行时间13 min,压力波动± <5 bar(标称工作压力约360 bar)。

序列(单日内进行的一系列测定运行)由重复的测定运行组成,即“标准品a、样品1、样品2、样品3、标准品b”等。这确保了样品中任何残留均可通过视觉识别(“鬼峰”的出现),并且色谱柱定期冲洗以维持其平衡(表现为基线稳定和/或基线噪声(BLN)不增加)。

MP在每个序列前新鲜配制,由水:乙腈(ACN):甲酸(FA)混合物组成,名义比例为96:4:0.1 v/v,配制方法为:在500 mL容量瓶中,将约300 mL水与20 mL ACN和500 µL 98% FA混合,然后用水定容至500 mL。环境温度为21 ± 5 °C。

GS-44的检测通过识别在4.6 ± 0.3 min洗脱的离散峰来实现,通过在240 ± 5 nm处测量光吸收进行定量(宽狭缝,5 Hz采样率);峰下面积(AUP)以吸收单位(mAU)由软件测定。序列开始时运行的两个标准品确认了GS-44峰的洗脱时间。

标准GS-44浓度(ng/mL)使用最佳拟合外标线性回归公式[4]由AUP计算,然后转换为µM(ng/mL/291.26 [GS-44的分子量])和样品浓度(µM × 样品稀释因子)。

2.6. 方法验证

该方法使用研究样品(n = 1034)、空白样品(已知未接受GS-44或REM的猫血浆或血清;n = 15)和标准品(n = 17)进行验证;依据国际人用药品注册技术协调会(ICH)制定的生物分析方法验证指南——生物分析方法全面验证和研究样品分析M10(2022年5月24日通过)(ICH M10)[35],对选择性、特异性、灵敏度、LLOQ、校准曲线(及线性)、准确度、精密度、基质效应、残留和稳定性进行了评估。

由于可用于校准品和QC样品基质的空白猫血浆体积有限,无法在整个研究期间使用该基质用于标准品和QC样品;因此,在验证后改用标准溶液替代。该决定得到了以下结果的支持:方法验证期间使用的标准溶液与空白血浆标准品结果相比具有等效性能。

选择使用外标法,因为对于这种简单的样品制备方法,使用内标预计不会带来明显益处[4,36]。

2.7. 选择性、特异性和样品处理

使用空白猫样品和研究猫样品(加标和不加标GS-44)评估选择性和特异性。将加标GS-44的空白猫样品中GS-44的保留时间与标准品和研究样品中获得的保留时间进行比较。

方法开发确定了确保GS-44峰出现在样品色谱图中不与研究样品中其他组分产生的峰共流出位置所需的HPLC条件。天然存在的物质以及所有可能的外源性化合物(包括兽用药物和畜主给予的化合物)种类繁多,无法逐一检测,因此该方法规定由一位作者(SWC)对每张色谱图进行实时检查,以确认GS-44峰未受污染。使用同时多通道数据采集评估峰纯度和无共洗脱杂质。在240 ± 5 nm、240 ± 10 nm、230 ± 10 nm和250 ± 10 nm处记录吸收。在GS-44峰的上升段、顶点和下降段目视比较这些波长和带宽下的洗脱图谱和相对吸光度比值,以确认光谱均一性并排除干扰性共洗脱的存在(图1)。此时,应无干扰峰(满足干扰峰<LLOQ响应20%的要求)。

[图:研究样本的色谱图,显示检测过程中记录的全部四个吸收通道。吸收波长 w]

论文配图

对于每个样品,两个稀释度在同一序列内测定,间隔至少三次其他测定运行。如果计算的GS-44浓度值差异超过15%,则重新运行一个或两个稀释度。如果重复测定不能纠正变异(通过识别离群值),则为该样品重新配制一对新鲜稀释液并测定。

两个稀释度作为样品制备或测定过程中任何误差(包括进样重现性)的自检。如果两个稀释度的结果彼此差异在15%以内,则样品测定结果报告为两个结果的平均值±1个标准差(SD)。

使用加标回收实验评估提取回收率。正常样品(n = 11,已知不含GS-44)和研究样品(n = 5,预先测定并含有已知量的GS-44)均加入GS-44溶液,当各取4 µL加入40 µL样品中时,最终浓度分别为2.86、5.71和11.42 µM。提取回收效率计算为观测浓度与预期浓度之间的百分比差异。

2.8. 校准曲线、值范围、检测上限(ULOD)和定量下限(LLOQ)

制备了17个标准溶液,范围为5043.39至2.59 ng/mL。

通过测定标准溶液并以mAU为单位的AUP对预期GS-44浓度(ng/mL)作图确定校准曲线。17个标准品各至少测定四次(范围4–21次)。以这些结果的简单线性回归模型确定校准曲线和相关系数。同样,为每个标准品计算累积SD,并对计算浓度作图,所得简单线性回归模型用于计算样品测定结果的适当SD [37]。

对于值范围,ULOQ设定为5045 ng/mL,相当于1:12稀释时血浆浓度为208 µM,该值大于临床病例的样品值(本系列中最大测定值为49.2 µM),也大于据报道在体外不引起细胞毒性效应的GS-44浓度(100 µM [3,6]);该值预计在临床病例中不会达到,但并非不可能。

LLOQ使用软件确定的BLN高度(mAU)计算,基于三个最低标准浓度的多次运行。LLOQ定义为GS-44峰高>5倍最大BLN值的标准浓度。

2.9. 精密度和准确度

通过分析每日序列内至少四个不同标准品(日内)和不同日序列中的相同标准品(日间)评估精密度和准确度。准确度表示为平均测定浓度与标示浓度的百分比偏差,精密度表示为变异系数(CV% = [标准差/均值] × 100)。可接受标准为准确度在85–115%以内(最低三个标准浓度为80–120%),精密度(CV)≤15%(最低三个标准品≤20%)。

2.10. 残留

残留发生在以下情况:(i)样品组分的保留时间大于运行时间,从而在后续运行中产生峰,或(ii)GS-44在一次运行中未从固相完全洗脱,因此在后续运行中释放并表现为未识别峰。通过观察连续运行评估这些情况,使用研究样品或最高浓度标准品GS-44,随后运行MP。

2.11. 稀释完整性

通过比较不同批次标准溶液的标准溶液测定结果与预期值评估标准溶液稀释完整性。

2.12. 稳定性

通过每日监测标准品结果来评估标准溶液的稳定性。若任一标准品的AUP与其新鲜配制标准品记录值偏差>15%(三个最低浓度标准品为>20%),则弃用该标准品。若该组标准品中>25%因此原因被排除,则弃用剩余标准品并重新配制一组。

通过比较每组新配制标准品的AUP与各先前标准品组的初始结果来监测储备液稳定性。

通过将样品在室温(RT)下放置4天、室温4天后冰箱温度4天或室温4天、冰箱温度4天后再冷冻4天后重复样品测定来评估样品稳定性。

2.13. 统计分析

使用Excel(Microsoft Corporation,Redmond,WA,USA)和GraphPad Prism v10.6.1 (892)(GraphPad Software,Boston,MA,USA)进行统计分析;p值<0.05视为具有统计学显著性。

3. 结果

3.1. 验证

选择性和特异性

样品、标准GS-44溶液和加标GS-44的空白猫样品的色谱图在4.6的洗脱时间产生GS-44峰,该时间在不同天数之间略有变化(±0.3 min),取决于MP配制时的温度(主要为ACN的体积效应)。

对于空白猫样品,基线接近或处于0.0 mAU,在2.5至9 min的基线时间内保持稳定,BLN<0.025 mAU,且在预期的GS-44洗脱时间4.6 ± 0.3 min无明显峰。对于加标样品,GS-44的AUP在预期值的<±20%范围内,未见内源性峰干扰GS-44峰。

空白猫血浆和研究样品的代表性色谱图示例见图2。对于空白猫血浆,没有与研究样品中GS-44占据相同洗脱时间的峰。向含有内源性GS-44的样品中加入GS-44未改变洗脱时间,也未产生峰前伸、拖尾、分裂、肩峰、基座增宽或保留时间变化。在另外11个研究样品上进行的加标回收实验观察到类似结果。这表明该测定具有选择性和特异性。

[图:(上)已知接受过 GS-441524(GS-44)的猫的研究样本色谱图,以及(下)空白猫样本的色谱图]

论文配图

3.2. 基质效应

GS-44峰未受样品中其他化合物的明显影响,通过对每张色谱图进行目视检查,在研究样品色谱图在四通道波长值和带宽下峰特征和洗脱时间均无变化,证明了这一点。

3.3. 校准曲线、数值范围、检测上限(ULOD)和定量下限(LLOQ)

校准数据(表1)最佳拟合简单线性回归模型,y = 0.00301x,截距设为0.0,r2 = 0.9974。也生成了二阶模型;然而,该模型未能正确拟合两个最低标准浓度结果,因此被弃用。

Table 1. Standard GS-441524 calibration dilutions and assay results, including percentage residual standard deviation (RSD%) and standard deviation (SD) in both ng/mL and µM.

Table 1 Standard GS-441524 calibration dilutions and assay results, including percentage residual standard deviation (RSD%) and standard deviation (SD) in both ng/mL and µM.
Standard Set UsedNumber of Standard AssaysExpected Standard Concentration ng/mLSD for Each Standard ng/mLMean for Each Standard ng/mLRSD% for Each StandardExpected Concentration µMSD for Each Standard µM
1135048.28259.685043.395.1517.3330.892
2213527.71195.783504.245.5912.1120.672
3181853.1678.091946.704.016.3630.268
411950.6538.17950.904.013.2640.131
511834.5336.65828.144.432.8650.126
621717.6538.92712.335.462.4640.134
718600.0011.28619.471.822.0600.039
817481.5826.73466.955.721.6530.092
917362.3823.97366.646.541.2440.082
1012242.3814.52240.936.030.8320.050
1113121.197.66126.336.060.4160.026
121960.603.7959.616.360.2080.013
13232.111.0130.853.270.1100.003
141416.001.5315.749.750.0550.005
15410.360.339.783.410.0360.001
1645.180.174.973.390.0180.001
17102.590.232.409.520.0090.001

SD数据(表1)最佳拟合y = 0.0521x − 0.0071,r2 = 0.9880。

BLN值从未超过0.026 mAU;该值用于计算LLOQ。

加标回收研究产生的平均回收率为95%,范围为86%至110%;绝对偏差为5%。由精密度实验计算得出的回收率为99%(范围89–105%),估计绝对偏差为1%。

根据标准品测定,LLOQ设定为2.59 ng/mL(表2),相当于1:6.5稀释时血浆浓度为0.06 µM。表2显示了该选择的依据,包括对较低三个标准稀释液的测定(n = 19)及其各自的吸光度、峰高和BLN值的计算。

Table 2. Lowest three GS-441524 standard assay results and LLOQ selection criteria. LLOQ 2.59 ng/mL, peak height &gt;5 × baseline noise (BLN) (Ratio height to BLN).

Table 2 Lowest three GS-441524 standard assay results and LLOQ selection criteria. LLOQ 2.59 ng/mL, peak height >5 × baseline noise (BLN) (Ratio height to BLN).
StandardGS-44 Conc, Expected; ng/mLGS-44 Assay Calculated; ng/mLAs% of ExpectedAUP; mAU/minHeight; mAUBLN Height; mAURatio Height to BLN
1510.369.6093%0.02890.3400.02116.190
1510.369.9796%0.03000.3400.01621.250
1510.367.4172%0.02230.2600.01123.636
1510.369.4091%0.02830.3300.01227.500
165.185.0598%0.01520.1700.01610.625
165.184.9596%0.01480.1720.01412.286
165.184.7291%0.01420.1700.01214.167
165.185.0898%0.01530.1700.01017.000
172.592.5398%0.00760.0900.0175.294
172.592.2687%0.00680.0800.0155.333
172.591.9374%0.00580.0800.0155.333
172.592.59100%0.00780.0700.0135.385
172.592.3691%0.00710.0800.0145.714
172.592.59100%0.00780.0800.0145.714
172.592.59100%0.00780.0900.0146.429
172.592.89112%0.00870.0900.0118.182
172.592.4996%0.00750.0900.0118.182
172.592.1382%0.00640.0800.0145.714
172.593.69142%0.01110.1300.0168.125

3.4. 精密度与准确度

除最低两个稀释度外,所有样品的真实度(以偏差表示)<5%,最低两个稀释度<10%;除最低两个稀释度外,精密度(以变异系数[CV]表示)<5%,最低两个稀释度<15%。

3.5. 残留

检查空白血浆或标准品及样品测定运行后立即运行的MP所获得的色谱图,在4.6 ± 0.3 min处均未显示任何峰,表明GS-44或其他样品组分在柱上的保留时间均未超过18 min;因此,在此运行时间和GS-44稀释范围内未发生残留。

3.6. 稳定性

储备液稳定性可接受,16个月内储备GS-44含量降低<5%。

标准溶液也非常稳定;批次间更换新鲜溶液的平均时间为43天。标准品组被弃用的最常见原因是 volume 因使用而耗尽。在此期间,观察到浓度略有增加(通常为1–3%),认为系使用过程中蒸发所致。

样品稳定性也可接受,在−20 °C(冷冻)、0–6 °C(冷藏)、20 °C(自动进样器)及室温条件下,10天内储存样品与新鲜测定样品结果之间的变异<11%。

3.7. 临床应用

共测定了来自361只猫的1034份样品。样品GS-44浓度(µM)结果被分配至24个一小时区间,代表给药后连续一小时时段的采样时间(例如,区间1 = 0–59 min,区间2 = 60–119 min等);以条形加须图绘制(图3)。黑色条表示数值范围,绿色条表示25%至75%百分位数,实心中间条为中位数。

[图:对于来自 361 只个体猫的 1034 份研究样本:y 轴——样本 GS-441524 浓度(µM),x 轴——样本]

论文配图

值得注意的是,观察到样品GS-44浓度存在较大变异,此已在别处报告和讨论[38]。

4. 讨论

本研究成功开发并验证了一种采用紫外检测的HPLC方法,用于定量接受该化合物治疗FIP的猫血样中的GS-44,并符合国际人用药品注册技术协调会M10文件《人用药品技术要求,生物分析方法全面验证和研究样品分析》[35]所规定的标准。

内源性(样品)GS-44和加标GS-44从已含GS-44的加标样品中洗脱,保留时间相同,表明该测定兼具选择性和特异性。

由于在研究样品和空白猫样品中均发现一种 unidentified 化合物(与GS-44无关)在11 min处产生洗脱峰,故选择13 min的运行时间以防止该化合物污染后续色谱图。

由于FIP临床表现和全身效应的复杂性,预期部分研究样品会含有外源性化合物,如皮质类固醇、抗生素、食欲刺激剂、镇痛药、镇静剂或其他抗病毒化合物[4];然而,若任何此类化合物既与GS-44完全相同的保留时间洗脱(如本方法所见)又在230至250 nm之间吸收光,则纯属巧合。操作人员在色谱图展开过程中逐一观察,可确保洗脱时间在GS-44值±0.3 min范围内的化合物不产生重叠峰及随后的GS-44峰伪影。未观察到此类情况。

所有研究样品中GS-44与猫血浆其他组分分离良好,且该方法使用简单的样品制备技术和直接的等度MP洗脱,产生了始终可重现的结果。

本文所述的HPLC方法也满足相对环保方法的要求[39];ACN是一种有毒物质(也称为甲基氰),需要使用分馏法进行专业回收(一种昂贵且能耗高的方法),或者(如本研究所用)使用氢氧化钠进行化学中和[40],因此希望流动相中ACN的浓度尽可能低。本研究方法每次运行使用0.78 mL ACN,而最近一种非MS的HPLC方法[4]每次运行使用5.39 mL ACN,另一种使用MS的方法[41]每次运行使用2.61 mL。使用HPLC-MS或HPLC-MS/MS的检测需要永久保持在真空状态的设备(持续消耗电能并产生废热和噪音),使用高纯度氮气、氩气或氦气(制造过程能耗高),需要使用高纯度溶剂和专用样品瓶(比LC等效耗材能耗更高),还可能需要使用放射性内标(生产能耗高)。开发使用独立HPLC的方法避免了对依赖此类环境成本高昂的复杂设备的依赖。

预计方法开发将继续使用UHPLC专用色谱柱以缩短运行时间和减少试剂用量,从而进一步降低检测的环境影响。

其他核苷类似物抗病毒药物如molnupiravir(EIDD-2801)已被用于治疗FIP[5],本文所述方法目前正在修改并用于支持该化合物的TDM,通过检测其母体药物N4-羟基胞苷(NHC,EIDD-1931),该药物也用于治疗。

需要观察每张色谱图的展开情况在一定程度上限制了商业应用,因为该方法不具有“无人值守”的性质,因此需要经过培训的操作人员每天在整个样品序列期间进行观察。

该方法生成TDM数据的临床应用似乎是合理的,因为本文收集的样品结果(研究截至2026年6月仍在继续)显示给药后每个一小时时间段内数值范围较宽(图3)。这种较大变异表明受FIP影响的猫体内进行的代谢过程也存在类似范围的差异,以及GS-44给药后吸收和/或消除的相应变异性。这并不意外,因为FIP至少有四种表型(包括表达一种以上表型的猫),并影响年龄、性别、品种、饲养管理和免疫能力各异的人群。这些变异可能转化为GS-44向其预期作用部位(单核细胞和巨噬细胞)呈递的效率以及固有的抗病毒效果欠佳的潜在可能,从而导致治疗失败或缓解后复发。作为一种临床工具,TDM非常适合研究和量化这些变异。

5. 结论

本研究描述了一种血浆和血清GS-44的检测方法,在成本效益、使用易获取的技术、方法简单环保以及易于实施方面满足了诊断实验室和研究实验室的实际需求。我们希望这将鼓励临床实验室(取决于其采用的方法)在FIP治疗期间更广泛地使用TDM来监测GS-44(或许还有类似药物)的组织浓度。这反过来将使建立足够大的数据集成为可能,这些数据集代表FIP临床表现中固有的所有变量,并构建有意义的群体PK模型,进一步完善该疾病的治疗。

致谢

感谢各诊所所有提供猫样品的人员。

作者贡献

概念构思,S.W.C.和R.H.;方法学,S.W.C.;验证,S.W.C.;形式分析,S.W.C.;调查研究,S.W.C.;资源,S.W.C.;数据管理,S.W.C.;写作——初稿准备,S.W.C.;写作——审阅与编辑,D.A.G.-M.和R.H.;可视化,S.W.C.;监督,S.W.C.;项目管理,S.W.C.;资金获取,S.W.C.。所有作者均已阅读并同意稿件的发表版本。

机构审查委员会声明

动物研究方案经英国爱丁堡大学兽医伦理审查委员会(VERC)批准;VERC 63.22;日期为2022年12月4日。

知情同意声明

不适用。

数据可用性声明

数据将在合理请求下向swc@microsampling-laboratory.co.uk提供。

利益冲突

作者Stephen W. Cooke受雇于The Microsampling Laboratory Ltd.公司。其余作者声明,本研究在不存在任何可被解释为潜在利益冲突的商业或财务关系的情况下进行。

资金声明

本研究由非营利组织The Microsampling Laboratory Ltd.全额资助。

脚注

[内容请查看原文]

参考文献

[内容请查看原文]

关联数据

数据可用性声明

数据将在合理请求下向swc@microsampling-laboratory.co.uk提供。

查看英文原文▾

1. Introduction

Remdesivir (REM) and GS-441524 (GS-44) are anti-viral agents that are safe and effective when used to treat cats suffering from the various forms of feline infectious peritonitis (FIP) [1,2]. Both drugs are now legally available for prescription by veterinary surgeons in many countries from licensed veterinary pharmaceutical compounding companies.

The treatment regimens being used for these drugs are based on data derived from studies involving in vitro cell cultures and healthy cats [3,4,5,6,7,8] as well as retrospective reports of results from cats treated with legally sourced GS-44 and/or REM [4,7,8,9,10,11,12,13,14,15,16,17,18,19,20] or unverifiable compounds sourced illegally [21,22,23,24,25]. To date, published pharmacokinetic (PK) data for these drugs being administered under veterinary supervision to cats with clinical FIP are sparse (n = 6 [PK data] n = 22 [single sample data]) [8] and there is a need for robust investigation into dosages appropriate for each clinical presentation of FIP [8]. Given the limited predictive PK/pharmacodynamic (PD) data available for clinical cases, accurate quantification of GS-44 concentrations is an essential alternative to inform evidence-based dosing and therapeutic monitoring.

Therapeutic Drug Monitoring (TDM) is used to monitor the plasma concentration of target drugs to optimise treatment [26]; this is primarily to prevent under- or over-dosage, but also to support investigations into causes of failure in response to medication, side effects or relapses from remission.

TDM of GS-44 requires a laboratory method readily available to clinicians which can report results both rapidly (<7 days) and cost-effectively. At present, high performance liquid chromatography combined with mass spectroscopy (HPLC-MS or HPLC-MS/MS [27,28,29,30]) is the method most used by researchers; however, this utilises expensive and complicated equipment, that needs to be run by specialist technicians, it uses significant amounts of environmentally sensitive resources and is not (June 2026) offered for clinical use.

The aim of this paper is to describe the validation of a simple, low cost, effective HPLC method without using MS and using sparing amounts of environmentally sensitive resources, which may find a place in commercial diagnostic laboratories and provide clinicians with access to this important resource.

We also report the GS-44 assay results of 1034 samples originating from 361 cats undergoing treatment for the various clinical presentations of FIP.

2. Materials and Methods

2.1. Sample Origin and Handling

Feline serum (spontaneous clotting or serum gel) and plasma samples (ethylenediaminetetraacetic acid [EDTA] and/or heparin anticoagulated) were obtained as residual diagnostic material from routine clinical submissions to the Easter Bush Pathology Laboratory, Royal (Dick) School of Veterinary Studies, University of Edinburgh, and as diagnostic samples originating from cats receiving legally prescribed REM or GS-44 for the management of FIP. The use of residual clinical samples for assay development and validation was approved by the Veterinary Ethical Review Committee (VERC) of the University of Edinburgh, UK (VERC 63.22; approved 4 December 2022). Following diagnostic testing, residual material was transferred to The Microsampling Laboratory for method development and validation.

2.2. Chemicals, Reagents and Expendables

All solvents (including ultra-pure water) and chemicals were of HPLC grade (filtered at 0.22 µm) and purchased from LaserChrom HPLC Laboratories, London, UK, except for GS-44 (CAS: 1191237-69-0, purity > 99%, stored at −20 °C in its original shipping container), which was purchased from MedChemTronica AB, Stockholm, Sweden.

2.3. Preparation of Stock and Standard Solutions

Methanol (100%) was used to prepare a 36,600 ng/mL stock solution of GS-44, which was stored at −20 °C. Seventeen standard solutions, ranging from 5048.28 to 2.59 ng/mL GS-44, were prepared from the stock solution diluted with the HPLC mobile phase (MP) described below. Methanol was not used as a diluent for the standards as, during method development, excessive perturbation of the chromatogram’s baseline and peak broadening/shouldering occurred when using methanol, artefacts not exhibited when using MP as the diluent; this appeared to be a column specific phenomenon. Standard solutions were prepared in 2 mL HPLC sample vials and stored between 0 °C and +6 °C (laboratory refrigerator) and briefly vortexed before each assay sequence. Standard solutions were assayed together with study samples and discarded if their calculated concentration deviated more than 15% (more than 20% for the three lowest standard concentrations) from the expected values. Fresh standard sets were prepared as necessary (approximately every 6 months).

2.4. Extraction and Sample Preparation

For this paper, the term “sample” refers to plasma from EDTA- or heparin-anticoagulated blood or serum.

The GS-44 blood-plasma partitioning ratio for cats is believed to be 1.2 (five other species returned similar values [31]); thus, for completely haemolysed samples, assay values were likely to be approximately 20% higher than those obtained if haemolysis was not present. For this study, if red blood cells could not be separated from haemolysed samples by centrifugation, i.e., complete haemolysis had occurred, these samples were excluded.

Molecules causing icterus and lipaemia in samples would not cause interference with the method [32,33], so these artefacts were not considered to be interfering factors.

Each sample was prepared at 2 dilutions by mixing 220 µL methanol (to remove proteins by coagulation) with (a) 20 µL of sample (1 in 12 dilution) and (b) 40 µL of sample (1 in 6.5 dilution) using a positive pressure pipettor (Gilson Microman; Gilson UK, Dunstable. UK) to prevent viscous samples causing measuring errors. After vortexing twice for 10 s with a 30 s interval and centrifugation at a relative centrifugal force of 7700 g for 6 min, 210 µL of the resulting supernatant was transferred to a fresh 1.5 mL vial and evaporated to dryness under vacuum (GeneVac EZ-2, circa 60 min at ≤40 °C). The residue was redissolved in 210 µL MP, using an ultrasonic bath (5 min at ≤40 °C) and vortexing, and then filtered through a 4 mm 0.22 µm nylon syringe filter into a 0.3 mL HPLC vial for analysis.

2.5. Instrumentation and Chromatographic Condition

A Dionex (Fisher Scientific UK Ltd., Loughborough, UK) UltiMate RS 3000 UHPLC system, consisting of a four-channel MP vacuum degasser, temperature-controlled auto-sampler (20 °C), twin pumps, column oven (column 40 °C; post-column 23 °C) and a four-channel UV diode array detector (DAD) was used. Instrument control, data collection and chromatogram analysis were performed by Chromeleon v 7.2 (Fisher Scientific UK Ltd., Loughborough, UK) software. The sample injection volume was 40 µL, with needle washing using MP between injections.

The GS-44 molecule is highly polar (XlogP3-AA = −1.4 [34]) and so requires a high aqueous content MP to increase its retention time on C18 solid-phase columns sufficiently to ensure adequate peak separation from other sample components. The Waters XSelect HSS T3 100Å, 2.5 µm, 4.6 mm × 150 mm column (Waters Ltd., Wilmslow, UK), was selected as its solid phase was designed to be used with an MP containing up to 100% water and, during method development, provided satisfactory separation of the GS-44 peak from other sample components which also absorbed UV wavelengths between 220 and 260 nm. Elution was isocratic with a 1.5 mL/min MP flowrate and run time of 13 min, and pressure ± <5 bar (nominal working pressure circa 360 bar).

Sequences (a series of assay runs performed on a single day) were made up of repeats of assay runs “standard a, sample 1, sample 2, sample 3, standard b”, etc. This ensured that any carry-over from a sample was visually identifiable (the presence of “ghost peaks”) and that the column was flushed on a regular basis to maintain its equilibrium (exhibited by baseline stability and/or no increase in baseline noise (BLN)).

The MP was made up freshly for each sequence and consisted of a water: acetonitrile (ACN): formic acid (FA) mixture, nominally 96:4:0.1 v/v, made by mixing, in a 500 mL volumetric flask, circa 300 mL water with 20 mL ACN and 500 µL 98% FA, and then increasing the volume up to 500 mL with water. Ambient temperature was 21 ± 5 °C.

Detection of GS-44 was achieved by identifying a discrete peak eluting at 4.6 ± 0.3 min, which was quantified by measuring light absorption at 240 ± 5 nm (wide slit, 5 Hz sampling rate); the area under the peak (AUP) in absorption units (mAU) was determined by the software. Two standards run at the start of a sequence confirmed the elution time for the GS-44 peak.

The standard GS-44 concentrations (ng/mL) were calculated from the AUP using the external standard linear regression formula of best fit [4] and then converted to µM (ng/mL/291.26 [molecular weight of GS-44]) and sample concentration (µM × sample dilution factor).

2.6. Method Validation

The method was validated using study samples (n = 1034), blank samples (plasma or serum from cats known as not receiving GS-44 or REM; n = 15) and standards (n = 17); it was assessed for selectivity, specificity, sensitivity, LLOQ, calibration curve (and linearity), accuracy, precision, matrix effect, carry-over and stability in accordance with the biological method validation guidelines prepared by the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use, Bioanalytical method full validation and study sample analysis, M10 adopted 24 May 2022 (ICH M10) [35].

Due to only small volumes of blank cat plasma being available for use as the matrix for the calibration and QC samples, it was not possible to use this matrix for standard and QC samples during the entirety of the study; instead, after validation, the standard solutions were substituted. This decision was supported by the equivalent performance of the standard solutions as used during method validation when compared with the blank plasma standard results.

Use of an external standard was selected as there was no likely benefit expected from using an internal standard with this uncomplicated sample preparation method [4,36].

2.7. Selectivity, Specificity and Sample Processing

Selectivity and specificity were assessed using both blank and study cat samples with and without being spiked with GS-44. The retention time of GS-44 with blank cat samples spiked with GS-44 was compared with that obtained for standards and study samples.

Method development determined the HPLC conditions required to ensure that the GS-44 peak occurred at a point in the sample chromatogram where peaks resulting from other components present in the study samples would not coincide. Naturally occurring substances as well as the gamut of possible exogenous compounds (including veterinary medications and owner-administered compounds) were too numerous to be individually tested, so the method stipulated real-time examination of each chromatogram by one author (SWC) to confirm that the GS-44 peak was unadulterated. Peak purity and the absence of co-eluting impurities were evaluated using simultaneous multi-channel data collection. Absorption was recorded at 240 ± 5 nm, 240 ± 10 nm, 230 ± 10 nm, and 250 ± 10 nm. The elution profiles and relative absorbance ratios across these wavelengths and bandwidths were visually compared at the upslope, apex, and downslope of the GS-44 peak to confirm spectral homogeneity and rule out the presence of interfering co-elutions (Figure 1). At this point, there would be no interfering peaks (satisfying the requirement for interfering peaks being <20% of the LLOQ response).

[图:Chromatogram of a study sample, showing all four absorption channels recorded during the assay. Absorption wavelengths w]

论文配图

For each sample, the two dilutions were assayed within the same sequence separated by at least three other assay runs. If the calculated GS-44 concentration values differed by more than 15%, then either one or both dilutions were re-run. If the repeats could not correct the variation (by identifying the outlier value), then a fresh dilution pair was made up for that sample and assayed.

The two dilutions acted as a self-check for any errors in sample preparation or the assay process (including injection reproducibility). If the results of the two dilutions fell within 15% of each other, the sample assay results were reported as the average of the two results ±1 one standard deviation (SD).

Extraction recovery was assessed using spike recovery experiments. Both normal samples (n = 11, known not to contain GS-44) and study samples (n = 5, pre-assayed and containing known amounts of GS-44) were spiked with solutions of GS-44 prepared to produce the final concentrations of 2.86, 5.71 and 11.42 µM when 4 µL of each was added to a 40 µL sample. Extraction recovery efficiency was calculated as the percentage difference between the observed and expected concentrations.

2.8. Calibration Curve, Range of Values, Upper Limit of Detection (ULOD) and Lower Limit of Quantification (LLOQ)

Seventeen standard solutions were prepared with a range of 5043.39 to 2.59 ng/mL.

The calibration curve was determined by assaying the standard solutions and plotting the AUP in mAU against the expected GS-44 concentration in ng/mL. Each of the 17 standards was assayed at least four times (range 4–21). A calibration curve and correlation coefficient were determined as a simple linear regression model of these results. Similarly, a cumulative SD was calculated for each standard and plotted against the calculated concentration and the resulting simple linear regression model was used to calculate an appropriate SD for the sample assay results [37].

For the range of values, the ULOQ was set at 5045 ng/mL, equivalent to a plasma concentration of 208 µM at a 1 in 12 dilution, which is a value greater than sample values from clinical cases (with a maximum assay value of 49.2 µM in our series) and greater than the GS-44 concentration reported to not cause cytotoxic effects in vitro (100 µM [3,6]); this is a value not anticipated to be achieved in clinical cases, but is not impossible.

The LLOQ was calculated using the BLN height (mAU) as determined by the software, over multiple runs of the three lowest standard concentrations. The LLOQ was defined as the standard concentration with a GS-44 peak height of >five times the maximum BLN value.

2.9. Precision and Accuracy

Precision and accuracy were evaluated by analysing at least four different standards within each day’s sequence (intra-day) and the same standards over different day’s sequences (inter-day). Accuracy was expressed as the percentage deviation of the mean measured concentration from the nominal concentration and precision as the coefficient of variation (CV% = [standard deviation/mean] × 100). Acceptable criteria were accuracy within 85–115% (80–120% for the lowest three standard concentrations) and precision (CV) ≤15% (≤20% for the lowest three standards).

2.10. Carry-Over

Carry-over occurs where either (i) a sample component had a retention time greater than the run time, thus producing a peak in the subsequent run, or (ii) where GS-44 was not completely eluted from the solid phase during a run, so it was released in the subsequent run and appeared as an unidentified peak. These were assessed by observing sequential runs, using either study samples or the highest concentration standard GS-44, followed by MP.

2.11. Dilution Integrity

Standard solution dilution integrity was assessed by comparing standard solution assay results with expected values for different batches of standard solutions.

2.12. Stability

The stability of the standard solutions was assessed by monitoring the standard results on each day. If the AUP of any standard differed by >15% (>20% for the three lowest standards) of that recorded for the freshly prepared standard, that standard was discarded. If >25% of the standards from that set were excluded for this reason, then the remainder were discarded and a new set prepared.

Stock solution stability was monitored by comparing the AUPs of each new set of standards with the initial results for each previous standard set.

Sample stability was assessed by repeating sample assays after samples had been held at room temperature (RT) for 4 days, RT 4 days then fridge temperature for 4 days or RT 4 days, fridge temperature 4 days then frozen for 4 days.

2.13. Statistical Analysis

Statistical analysis was carried out using Excel (Microsoft Corporation, Redmond, WA, USA) and GraphPad Prism v10.6.1 (892) (GraphPad Software, Boston, MA, USA); p-values <0.05 were considered statistically significant.

3. Results

3.1. Validations

Selectivity and Specificity

Chromatograms of sample, standard GS-44 solutions and GS-44 spiked blank cat samples produced GS-44 peaks at an elution time of 4.6, which varied slightly (±0.3 min) between days, depending on the temperature at which the MP was prepared (volumetric effect of, mainly, ACN).

For blank cat samples, the baseline was near to or at 0.0 mAU, stable for the baseline time between 2.5 and 9 min, exhibited a BLN of <0.025 mAU and had no peaks evident at the expected GS-44 elution time 4.6 ± 0.3 min. For spiked samples, the GS-44 AUP corresponded to <±20% of the expected value and no endogenous peaks were seen to interfere with the GS-44 peak.

Representative examples of chromatograms of blank cat plasma and a study sample are shown in Figure 2. For the blank cat plasma there are no peaks that occupy the same elution time as the GS-44 in the study sample. Spiking GS-44 into a sample containing endogenous GS-44 did not change the elution time, and did not produce peak fronting, tailing, splitting, shouldering, base broadening or retention time variation. Similar results were observed in the spike recovery experiments performed on another 11 study samples. This shows that the assay is selective and specific.

[图:Chromatogram of ( top ) a study cat sample from a cat known to have received GS-441524 (GS-44) and ( bottom ) a blank ca]

论文配图

3.2. Matrix Effect

The GS-44 peaks were not overtly affected by other compounds present in the samples, as demonstrated by the absence of changes in peak characteristics and elution times across the study sample chromatograms at the four-channel wavelength values and bandwidth, by visual inspection of each chromatogram.

3.3. Calibration Curve, Range of Values, Upper Limit of Detection (ULOD) and Lower Limit of Quantification (LLOQ)

The calibration data (Table 1) best fitted to a simple linear regression model, y = 0.00301x, with the intercept set to 0.0, r2 = 0.9974. A second-order model was also generated; however, this failed to correctly fit the two lowest standard concentration results and so was discarded.

Table 1. Standard GS-441524 calibration dilutions and assay results, including percentage residual standard deviation (RSD%) and standard deviation (SD) in both ng/mL and µM.

Table 1 Standard GS-441524 calibration dilutions and assay results, including percentage residual standard deviation (RSD%) and standard deviation (SD) in both ng/mL and µM.
Standard Set UsedNumber of Standard AssaysExpected Standard Concentration ng/mLSD for Each Standard ng/mLMean for Each Standard ng/mLRSD% for Each StandardExpected Concentration µMSD for Each Standard µM
1135048.28259.685043.395.1517.3330.892
2213527.71195.783504.245.5912.1120.672
3181853.1678.091946.704.016.3630.268
411950.6538.17950.904.013.2640.131
511834.5336.65828.144.432.8650.126
621717.6538.92712.335.462.4640.134
718600.0011.28619.471.822.0600.039
817481.5826.73466.955.721.6530.092
917362.3823.97366.646.541.2440.082
1012242.3814.52240.936.030.8320.050
1113121.197.66126.336.060.4160.026
121960.603.7959.616.360.2080.013
13232.111.0130.853.270.1100.003
141416.001.5315.749.750.0550.005
15410.360.339.783.410.0360.001
1645.180.174.973.390.0180.001
17102.590.232.409.520.0090.001

The SD data (Table 1) best fitted y = 0.0521x − 0.0071 and r2 = 0.9880.

Values for BLN never exceeded 0.026 mAU; this value was used to calculate LLOQ.

The spike recovery study produced an average recovery of 95%, ranging from 86 to 110%; the absolute bias was 5%. The recovery calculated from the precision experiment was 99% (range 89–105%) with an estimated absolute bias of 1%.

From the standard assays, the LLOQ was set at 2.59 ng/mL (Table 2), equivalent to a plasma concentration of 0.06 µM at a 1 in 6.5 dilution. Table 2 shows the justification for this choice with calculations for assays (n = 19) of the lower three standard dilutions and their respective absorption, peak height and BLN values.

Table 2. Lowest three GS-441524 standard assay results and LLOQ selection criteria. LLOQ 2.59 ng/mL, peak height &gt;5 × baseline noise (BLN) (Ratio height to BLN).

Table 2 Lowest three GS-441524 standard assay results and LLOQ selection criteria. LLOQ 2.59 ng/mL, peak height >5 × baseline noise (BLN) (Ratio height to BLN).
StandardGS-44 Conc, Expected; ng/mLGS-44 Assay Calculated; ng/mLAs% of ExpectedAUP; mAU/minHeight; mAUBLN Height; mAURatio Height to BLN
1510.369.6093%0.02890.3400.02116.190
1510.369.9796%0.03000.3400.01621.250
1510.367.4172%0.02230.2600.01123.636
1510.369.4091%0.02830.3300.01227.500
165.185.0598%0.01520.1700.01610.625
165.184.9596%0.01480.1720.01412.286
165.184.7291%0.01420.1700.01214.167
165.185.0898%0.01530.1700.01017.000
172.592.5398%0.00760.0900.0175.294
172.592.2687%0.00680.0800.0155.333
172.591.9374%0.00580.0800.0155.333
172.592.59100%0.00780.0700.0135.385
172.592.3691%0.00710.0800.0145.714
172.592.59100%0.00780.0800.0145.714
172.592.59100%0.00780.0900.0146.429
172.592.89112%0.00870.0900.0118.182
172.592.4996%0.00750.0900.0118.182
172.592.1382%0.00640.0800.0145.714
172.593.69142%0.01110.1300.0168.125

3.4. Precision and Accuracy

Trueness (as bias) was <5% for all except the lower two dilutions, which were <10%, and precision (as coefficient of variation [CV]) was <5% except for the lower two dilutions, which were <15%.

3.5. Carry-Over

Examination of the chromatograms obtained using blank plasma or an MP run immediately after both standard and sample assay runs did not show any peaks at 4.6 ± 0.3 min, indicating that neither GS-44 nor other sample components were retained by the column for more than 18 m; therefore, carry-over did not occur with this run time and GS-44 dilution range.

3.6. Stability

Stock solution stability was acceptable, with a reduction in stock GS-44 content <5% over a 16-month period.

Standard solutions were also very stable; the average time between batches being replaced with fresh ones was 43 days. The commonest reason for standard sets being discarded was because their volume was depleted from use. During that time, a slight increase in concentration was noted (typically 1–3%) and was believed to be due to evaporation during use.

Sample stability was also acceptable, with variation between stored and fresh assay sample results <11% over a 10-day period at temperatures ranging from −20 °C (freezer), 0–6 °C (fridge), 20 °C (autosampler) and ambient.

3.7. Clinical Application

A total of 1034 samples originating from 361 individual cats were assayed. The results of sample GS-44 concentration (µM) were assigned into 24 one-hour bins, representing the sample time after dose administration for sequential one-hour periods (e.g., bin 1 = 0–59 min, bin 2 = 60–119 min etc.); these were plotted as a bar and whisker plot (Figure 3). The black bar indicates the range of values, the green bar indicates the 25% to 75% percentiles with a solid intermediate bar being the median value.

[图:For 1034 study samples originating from 361 individual cats: y-axis—sample GS-441524 concentration (µM) for (x-axis) sam]

论文配图

Of note is the fact that a wide variation in sample GS-44 concentration was observed, and this is reported and discussed elsewhere [38].

4. Discussion

This study successfully developed and validated an HPLC method using ultraviolet detection for the quantification of GS-44 in blood samples from cats being treated with the compound for FIP, and satisfied the standards set out by the International Council for Harmonisation in the document M10, Technical Requirements for Pharmaceuticals for Human Use, Bioanalytical method full validation and study sample analysis [35].

Both endogenous (sample) GS-44 and spike GS-44 were eluted from spiked samples already containing GS-44 with the same retention time, indicating that the assay is both selective and specific.

Because an unidentified compound (unrelated to GS-44), found in both study and blank cat samples, produced a peak eluting at 11 min, a run time of 13 min was chosen to prevent this compound from contaminating subsequent chromatograms.

Due to the complexity of FIP’s clinical presentation and systemic effects it would be expected for some of the study samples to contain exogenous compounds such as corticosteroids, antibiotics, appetite stimulants, analgesics, sedatives or other antiviral compounds [4]; however, it would be singularly coincidental if any such compounds both eluted with exactly the same retention time as GS-44 (as found with this method) and absorbed light between 230 and 250 nm. Observing each chromatogram as it developed allowed the operator to ensure that no compounds with elution times ± 0.3 min of the GS-44 value produced overlapping peaks and subsequent GS-44 peak artefacts. None were observed.

There was very good separation of GS-44 from other components of cat plasma noted for all the study samples and the method produced consistently reproducible results using a simple sample preparation technique and straight-forward isocratic MP elution.

The HPLC method described here also satisfies the requirements of a, comparatively, eco-friendly method [39]; ACN is a toxic substance (also designated methyl cyanide) that requires either professional recycling using fractional distillation (an expensive and energy intensive method) or (as used in this study) chemical neutralisation using sodium hydroxide [40], so it was desirable to have the smallest possible concentration of ACN in the MP. The study method uses 0.78 mL ACN per run, whilst one recent non-MS HPLC method [4] used 5.39 mL ACN per run and another, using MS [41], used 2.61 mL per run. Assays utilising HPLC-MS or HPLC-MS/MS require equipment that has to be maintained at vacuum permanently (using electrical energy and producing waste heat and noise 24/7), uses high purity nitrogen, argon or helium (which are energy demanding to manufacture), requires the use of high purity solvents and specialized sample vials (more energy demanding than LC equivalents) and may also require radioactive internal standards (energy demanding to produce). Developing a method using stand-alone HPLC obviates the need to use equipment dependent on such environmentally costly complexities.

It is anticipated that method development will continue with the use of UHPLC-specific columns to reduce run times and reagent usage, thus further reducing the environmental impact of the assay.

Other nucleoside analogue antiviral drugs such as molnupiravir (EIDD-2801) have been used to treat FIP [5] and the method described in this paper is currently being modified and used to support TDM for this compound by assaying its parent drug, N4-hydroxycytidine (NHC, EIDD-1931), which is also used therapeutically.

The requirement of observing the development of each chromatograph restricts the commercial use slightly, as the method is not of a “walk-away” nature, so trained operators would be required to observe throughout the sample sequence each day.

The clinical application of the method to generate TDM data, appears to be justified as the sample results collected for this paper (the study continues as of June 2026) show a wide range of values for each one-hour time bin after drug administration (Figure 3). This wide variation suggests a similarly wide range of metabolic processes being carried out in cats affected by FIP and the consequent variability with which GS-44 is absorbed and/or eliminated after administration. This is not unexpected as FIP has at least four phenotypes (including cats expressing more than one) and affects a disparate population of age, sex, breed, husbandry and immunological competence. Such variations may be translated into how efficiently GS-44 is presented to its intended site of action (monocytes and macrophages) and the inherent potential for sub-optimal antiviral effects leading to failed treatments or relapse from remission. As a clinical tool, TDM is well-suited to investigate and quantify these variations.

5. Conclusions

The study describes an assay method for plasma and serum GS-44 which satisfies the practical requirements of both diagnostic and research laboratories in respect of cost effectiveness, usage of readily accessible technology, simple and eco-friendly method and ease of implementation. We hope that this will encourage, depending on the method being adopted by clinical laboratories, a more widespread use of TDM for monitoring the tissue concentrations of GS-44 (and perhaps similar drugs) during the treatment of FIP. This, in turn, will make it possible to populate sufficiently large data sets representing all the variables inherent in the clinical expression of FIP and construct meaningful population PK models, further refining the treatment of this disease.

Acknowledgments

Thank you to everyone in the clinics who provided samples from the cats.

Author Contributions

Conceptualization, S.W.C. and R.H.; methodology, S.W.C.; validation, S.W.C.; formal analysis, S.W.C.; investigation, S.W.C.; resources, S.W.C.; data curation, S.W.C.; writing—original draft preparation, S.W.C.; writing—review and editing, D.A.G.-M. and R.H.; visualization, S.W.C.; supervision, S.W.C.; project administration, S.W.C.; funding acquisition, S.W.C. All authors have read and agreed to the published version of the manuscript.

Institutional Review Board Statement

The animal study protocol was approved by the Veterinary Ethical Review Committee (VERC) of the University of Edinburgh, UK; VERC 63.22; dated 4 December 2022.

Informed Consent Statement

Not applicable.

Data Availability Statement

Data will be made available on reasonable request to swc@microsampling-laboratory.co.uk.

Conflicts of Interest

Author Stephen W. Cooke was employed by the company The Microsampling Laboratory Ltd. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Funding Statement

This research was funded in full by The Microsampling Laboratory Ltd., a not-for-profit organization.

Footnotes

[内容请查看原文]

References

[内容请查看原文]

Associated Data

Data Availability Statement

Data will be made available on reasonable request to swc@microsampling-laboratory.co.uk.

本文为开放获取(Open Access)文章,全文转载自 Europe PMC,遵循原许可协议(cc by); 来源:Pathogens (Basel, Switzerland) · DOI: 10.3390/pathogens15090913,版权归原作者及出版方所有;中文为机器翻译,临床术语以英文原文为准。 查看原文 ↗

跳转原文 ↗ 返回列表

快捷键 S 一键收藏本文,收藏内容在 我的收藏 查看;PubMed 位于境外,直接访问原文可能需要国际网络环境,站内摘要无需翻墙即可阅读。